Supplementary Fig. 2: Assembly of smoke exhaust and smoke containment modules.
From: Biomimetic smoking robot for in vitro inhalation exposure compatible with microfluidic organ chips

a. The Smoke Exhaust Module is assembled by: i) mounting 1”x1” fans to mounts using m2 screws, ii) inserting bulkhead mount barbs into barb mounts, iii) acrylic solvent bonding the remaining acrylic walls (back, top, outer_1, outer_2, and inner). b. The Smoke Containment Module Doors (2) are assembled by acrylic solvent bonding: i) door handles to each acrylic front door and back door, ii) rail_1 and rail_2 parts to each other, iii) rail assembly to side wall. c. The Smoke Containment Module is assembled by acrylic solvent bonding Sidewalls (2), Front wall, Back wall, and Top wall, making sure walls are facing the correct orientations as shown in figure. d. The Smoke Exhaust and Containment are assembled by acrylic solvent bonding exhaust module to top wall of the Smoke Containment Module, making sure exhaust module is facing the correct orientation as shown in figure.