Table 1 Metrological parameters (regression and validation data) of the techniques employed for distigmine determination.
Parameter | ISE | UV | HPLC | TLC |
---|---|---|---|---|
Accuracy (mean ± S.D.)a | 100.51 ± 0.531 | 100.53 ± 0.482 | 100.19 ± 1.344 | 100.62 ± 0.791 |
Specificity | 100.49 ± 1.486 | 100.78 ± 0.662 | 99.93 ± 1.424 | 100.42 ± 1.686 |
Repeatabilityb | 0.671 | 0.971 | 1.197 | 0.783 |
Intermediate precisionc | 0.771 | 1.087 | 1.234 | 0.962 |
Sloped | 27.51 | 0.0196 | 0.0619 | 0.2231 |
Intercept | 309.6 | 0.0095 | 0.06 | 0.0101 |
Correlation coefficient | 0.9993 | 0.9999 | 0.9998 | 0.9998 |
Range | 0.15 μM–10.0 mM | 17.35–104.10 μM | 6.94–55.52 μM | 2–12 μg/spot |
LODe | 0.078 μM | 5.22 μM | 1.65 μM | 0.62 μg/spot |
LOQf | 0.099 μM | 15.82 μM | 5.01 μM | 1.88 μg/spot |