Figure 2 | Scientific Reports

Figure 2

From: Measurement of subcutaneous fat tissue: reliability and comparison of caliper and ultrasound via systematic body mapping

Figure 2

Bland–Altman-Plot (a): intrarater caliper reliability (mean:-0.01 ± 0.02 cm; 95% limits of agreement − 0.04 to 0.05 cm); (b): intrarater US reliability (mean: 0.00 ± 0.03 cm; 95% limits of agreement of − 0.062 to 0.07 cm); (c): interrater caliper reliability (mean: 0.06 ± 0.11 cm; 95% limits of agreement of − 0.16 to 0.28 cm); (d): interrater US reliability (− 0.01 ± 0.15 cm; 95% limits of agreement of − 0.30 to 0.29 cm); SFT, subcutaneous fat tissue.