Table 1 Specifications of Octavius 729, Octavius 1500, and MatriXX ion chamber detector arrays.

From: Application of a linear interpolation algorithm in radiation therapy dosimetry for 3D dose point acquisition

Characteristics

Octavius 729

Octavius 1500

MatriXX

Detector type

Vented parallel plate ion chamber

Vented parallel plate ion chamber

Vented parallel plate ion chamber

Detector number

729

1405

1020

Detector diameter

5 mm

4.4 mm

4.4 mm

Detector shape

Square

Rectangle

cylindrical

Detector height

5 mm

3 mm

5.5 mm

Detector volume

0.125 cm3

0.058 cm3

0.07 cm3

Centre spacing

10 mm

7.07 mm

7.62 mm

Maximum dose rate

12 Gy/min

12 Gy/min

12 Gy/min

Working voltage

400 V

400 V

500 V

Maximum measuring area

27 × 27 cm2

27 × 27 cm2

24 × 24 cm2

Detector shape

Cuboid

Cuboid

Cylindrical

Absorber material on top of the array

5 mm polystyrene

5 mm polystyrene

3 mm ABS resin