Table 1 Metrological parameters (regression and validation data) of the techniques employed for distigmine determination.

From: A green compliant hand-held selective electrode device for monitoring active pharmaceuticals and the kinetics of their degradation

Parameter

ISE

UV

HPLC

TLC

Accuracy (mean ± S.D.)a

100.51 ± 0.531

100.53 ± 0.482

100.19 ± 1.344

100.62 ± 0.791

Specificity

100.49 ± 1.486

100.78 ± 0.662

99.93 ± 1.424

100.42 ± 1.686

Repeatabilityb

0.671

0.971

1.197

0.783

Intermediate precisionc

0.771

1.087

1.234

0.962

Sloped

27.51

0.0196

0.0619

0.2231

Intercept

309.6

0.0095

0.06

0.0101

Correlation coefficient

0.9993

0.9999

0.9998

0.9998

Range

0.15 μM–10.0 mM

17.35–104.10 μM

6.94–55.52 μM

2–12 μg/spot

LODe

0.078 μM

5.22 μM

1.65 μM

0.62 μg/spot

LOQf

0.099 μM

15.82 μM

5.01 μM

1.88 μg/spot

  1. aThe accuracy (n = 3), average of three concentrations (1.0 × 10–3, 1.0 × 10–4 and 1.0 × 10–5 M for ISE), (15.0, 35.0 and 55.0 µg/mL for UV), (10.0, 22.0 and 30.0 µg/mL for HPLC) and (5.0, 7.0 and 9.0 µg/spot for TLC).
  2. bThe intra-day (n = 3), RSD% of concentrations (5.0 × 10–3, 6.0 × 10–4 and 8.0 × 10–5 M for ISE), (20.0, 30.0 and 50.0 µg/mL for UV), (8.0, 16.0 and 24.0 µg/mL for HPLC) and (2.0, 6.0 and 10.0 µg/spot for TLC).
  3. cThe inter-day (n = 3), RSD% of concentrations (5.0 × 10–3, 6.0 × 10–4 and 8.0 × 10–5 M for ISE), (20.0, 30.0 and 50.0 µg/mL for UV), (8.0, 16.0 and 24.0 µg/mL for HPLC) and (2.0, 6.0 and 10.0 µg/spot for TLC).
  4. dAverage of three determinations.
  5. eLimit of detection for ISE is measured according to IUPAC by interception of the extrapolated arms of nonresponsive and Nernstian segments of the calibration plot of Figure S7. While for other methods, LOD is calculated according to ICH from the standard deviation (σ) of the regression residuals and the slope of the calibration curve (S) according to the following equations: LOD = 3.3(σ/S).
  6. fLOQ is calculated according to ICH from the standard deviation (σ) of the regression residuals and the slope of the calibration curve (S) according to the following equations: LOQ = 10(σ/S).