Abstract
mRNA measurement is dominated by RT-PCR, which requires expensive laboratory equipment and personnel with advanced degrees. Loop-mediated isothermal amplification (LAMP) is a versatile technique for detecting target DNA and RNA. The sensitivity of LAMP in early reports has been below that of the standard RT-PCR tests. Here, we report the use of a fluorescence-based RT-LAMP protocol to measure CDX2 expression patterns, which match extremely well to the standards of sophisticated RT-PCR techniques (r = 0.99, p < 0.001). The assay works on diverse sample types such as cDNA, mRNA, and direct tissue sample testing in 25 min compared to more than 3 h for RT-PCR. We have developed a new protocol for designing RT-LAMP primers that reduce false positives due to self-amplification and improve quantification. A simple device with a 3D-printed box enables the measurement of mRNA expression at home, outdoors, and point-of-care setting.
Similar content being viewed by others
Introduction
Loop-mediated isothermal amplification (LAMP) protocols1 have become a prevalent method of developing diagnostic assays2,3,4,5,6,7,8,9,10,11,12,13,14. Diagnostics based on LAMP amplification are a compelling alternative to polymerase chain reaction (PCR) because LAMP can be performed without commercial thermocyclers, significantly reducing cost and time to result1,15. In addition, LAMP has advantages over PCR for targeting sequences due to its robustness against inhibitors16,17 and its high specificity, using four to six primers that identify six to eight regions on the template for amplification1. The simplicity of isothermal amplification enables translation to a simple point-of-care device based on disposable cartridges9,10,18. LAMP assay kits are commercially available, but almost all of them use LAMP for detection as opposed to reliable quantification directly from the sample11,12,19,20,21,22,23,24,25,26,27. There have been some attempts at reliable quantification using Reverse transcription- Loop-mediated isothermal amplification (RT-LAMP), which has advantages over Reverse transcription-polymerase chain reaction (RT-PCR), such as simpler instrumentation and faster reaction times7,28,29. However, it has gained less popularity for quantification as RT-PCR. One reason is that RT-PCR has been extensively validated and is widely accepted as a gold standard for quantification of RNA targets, which leads researchers and regulatory agencies to prefer RT-PCR for quantification30,31. Additionally, RT-LAMP is relatively new compared to RT-PCR; thus, there may be less familiarity with the technique and fewer established protocols. Besides, RT-LAMP may have some limitations in terms of specificity and sensitivity compared to RT-PCR, which could make quantification more challenging11,32.
Given all the above points, ongoing efforts are still to improve and standardize RT-LAMP for quantification purposes. As with any new technology, it may take time for RT-LAMP to gain wider acceptance and for researchers to become more comfortable with its use for quantification. Our goal is to evaluate if RT-LAMP can be used to measure gene expression values robustly and reliably and to address current challenges for RT-LAMP technologies, which include complex primer design steps and self-amplification of the primers that lead to false positive results. Here, we have developed a new protocol that improves previous primer design steps, and we estimate the time from fresh tissue harvesting to signal detection for gene expression quantification for two genes: caudal type homeobox transcription factor 2 (CDX2) and Actin beta (ACTB) using RT-LAMP. CDX2 is a master regulator of intestinal development and oncogenesis33,34; its expression is highly specific to the intestinal epithelium35,36,37 and is used as a diagnostic biomarker for colorectal cancer38. ACTB is a popular housekeeping genes that is highly expressed in almost all cells39. The methods proposed in this manuscript can be readily generalizable to many other genes.
Results and discussion
Overview of RT-LAMP assays and comparative experimental design with RT-PCR
We compared the RT-LAMP based approach directly with RT-PCR for measuring CDX2 and ACTB gene expression values (Fig. 1, Fig. S1). Schematic experimental designs to test RT-PCR and RT-LAMP data are presented in Fig. S1. We measured the time from sample collection to the fluorescent signal crossing the detection threshold for both genes in\ RT-PCR and RT-LAMP settings (Fig. 1A, Fig. S1). We also tested multiple tissue samples in human and mouse for tissue specific gene expression patterns (Fig. S1A). CDX2 is known to be colon tissue-specific, matched against RT-PCR and RT-LAMP data. Additionally, we tested diverse sample types, such as cDNA, mRNA, and Tissue QuickExtract, in both settings (Fig. S1B). We performed serial dilution experiment to check quantitative features for both RT-PCR and RT-LAMP data. We compared the Ct values using a correlation test (Fig. S1C). Our goal in this experiment was to evaluate whether RT-LAMP data can be used reliably to measure gene expression. We also tested if the RT-LAMP assay can be performed at home by building a 3D-printed device (Fig. 1B).
CDX2 RT-LAMP expression is specific to colon tissue
All four primers for the human gene ACTB and CDX2 were tested using RT-LAMP protocol (Fig. S2, Fig. 2) in four different human tissue cDNA samples (1 × Lung, 2 × Colon, 1 × Blood; Fig. 2C-i, iv) to check tissue-specific expression patterns. As expected, cDNA was amplified in both ACTB-p2 (Fig. 2B-ii) and ACTB-p1 (Fig. 2B-iii) primer sets. CDX2 cDNA was amplified only in the two colon samples used for CDX2-p2 (Fig. 2B-v) and CDX2-p1 primer sets (Fig. 2B-vi). Our data agree with previous findings that LAMP is highly specific1,15,23, and CDX2 is expressed specifically in the colon tissue35,36,37.
Rapid CDX2 expression measurement using RT-LAMP and comparison with RT-PCR. (A) Five parameter logistic (5PL) model for comparative analysis of RT-PCR and RT-LAMP data. 5PL model parameters: baseline (a), slope factor (b), inflection point (c), plateau (d), and asymmetry factor € are displayed for a sample RT-PCR (left) and RT-LAMP (right) data. The X-axis displays time in minutes, and the y-axis displays the raw fluorescence intensity. 5PL model is computed using least squares regression, and the model predicted data is shown as a blue line on top of the original raw data as grey dots. The inflection point is shown using a vertical red line. (B) RT-LAMP experiments using hACTB, ACTB2, CDX2, and hCDX2 LAMP primer sets on three different human tissues (lung, colon and blood) cDNA samples. (C) RT-LAMP experiment is performed on cDNA, mRNA, and Tissue quick extract samples. Welch's two sample two-tailed unpaired t-test is performed to compute the p values.
RT-LAMP can be performed directly from cDNA, mRNA and tissue QuickExtract
To test the versatility of RT-LAMP, three different sample types were used: cDNA, mRNA, and tissue QuickExtract. RT-LAMP using ACTB-p2 primer set was able to amplify the target sequence using both mRNA and cDNA samples from human Blood (Fig. 2C-ii). Similarly, Cdx2-p2 primer set was able to amplify DNA using both mRNA and cDNA samples from mouse colon tissue (Fig. 2vC-iii). Next, we tested if gene expression can be measured directly from fresh tissue samples processed using Lucigen QuickExtract solution. Mouse blood was processed using Lucigen QuickExtract solution, and RT-LAMP protocol using Actb-p1 primer set was able to amplify Actb mRNA directly from the QuickExtract solution (Fig. 2C-iv). To validate this further, mouse colon, lung, kidney, spleen, and blood samples were used to test on all four mouse primer sets (Actb-p1, Actb-p2, Cdx2-p1, Cdx2-p2; Fig. S3A-i). As expected, Actb-p1 (Fig. S3A-ii) and Actb-p2 (Fig. S3A-iii) amplified all samples, while only the colon sample was amplified using Cdx2-p1 (Fig. S3A-iv) and Cdx2-p2 (Fig. S3A-v) primer set.
RT-LAMP based gene expression matches well with RT-PCR
To check the accuracy of gene expression measurement using RT-LAMP, two independent serial dilutions of human cDNA samples were prepared and processed using both RT-LAMP and RT-PCR protocols (Fig. 3A-i). For RT-LAMP protocol CDX2-p2 and ACTB-p2 primer sets were used, while RT-PCR protocol used only the F3 and B3 primers from the CDX2-p2 and ACTB-p2 primer sets. Both RT-PCR and RT-LAMP data were modeled using the five-parameters logistic (5PL) function, and all parameters were estimated using logistic regression40,41. The inflection points for RT-PCR and RT-LAMP were highly correlated for both CDX2-p2 (r = 0.99, p = 1.27e-05, Fig. 3A-ii) and ACTB-p2 (r = 0.97, p = 0.000271, Fig. 3A-ii). Another independent serial dilution experiment showed similar results for both CDX2-p2 (r = 0.97, p = 4.03e-05, Fig. 3A-iii) and ACTB-p2 (r = 0.98, p = 1.94e −05, Fig. 3A-iii). The results in cDNA quantification need to be repeated using mRNA quantification to check the overall performance of the RT-LAMP assay compared to RT-PCR in the future. Our results suggest that CDX2 expression can be reliably measured using RT-LAMP. While in our assay, RT-LAMP was faster than RT-PCR (Fig. S4), it may not be generalizable as the optimization in both settings may lead to variable results. The quantification results may be generalizable to fundamental nucleic acid detection using RT-LAMP techniques.
Correlation between RT-PCR and RT-LAMP. (A) Serial dilution was used to prepare human colon cDNA samples in various concentrations. hACTB and hCDX2 LAMP primer sets were used to perform RT-LAMP experiments from the diluted samples. F3 and B3 primers (hACTB and hCDX2) were used to perform RT-PCR experiments on the same diluted samples. The 5PL model was used to identify the inflection point for RT-LAMP and RT-PCR data and visualized using a scatter plot. This experiment was repeated in two human individuals’ cDNA samples (left and right). Correlation tests between RT-PCR and RT-LAMP data for both primers were calculated and displayed as scatter plots using Python seaborn lmplots with the p-values. The confidence interval around the regression line is indicated with shades. (B) 3D-model (4’’ × 3’’ × 5’’) of the 3D-printed box and device assembly details for use at home and outdoor settings. It uses an Arduino Pro Mini microcontroller and ESP32-CAM camera module to detect fluorescence signals. Results are displayed in a 0.96" 128X64 OLED LCD Display. Data from the device is collected over WIFI using an ESP32 module to a cell phone or a computer.
mRNA expression can be performed at home
To demonstrate if mRNA expression measurement can be performed outside a laboratory setting, we build a simple and affordable device (length = 4 in, width = 3 in, height = 5 in) that can be built using commercially available components (Fig. 3B). The camera can be easily replaced with a cell phone camera which can collect, analyze, and share data efficiently. CDX2-p2 RT-LAMP primers successfully quantified mRNA data at home (Fig. S5A), in laboratory benches (Fig. S5B), and in outdoor settings (Fig. S5C). This work demonstrates that mRNA expression measurement will not be limited to laboratory settings only. Consumer access to this assay and data, enabling innovative healthcare and hygiene applications. This technology will significantly reduce the expertise of the personnel needed to operate the device and quantify mRNA expression.
Modified RT-LAMP primer design steps to improve quantification and reduce false positive
One of the major challenges for the RT-LAMP is the complex primer design steps, which frequently produce primers that can self-amplify given enough time without any input sample42. Self-amplification will lead to false positives in the result. Efficient quantification of RT-LAMP technologies depends on quick amplification of input nucleic acid and delayed or no self-amplification. We found a new relationship between efficient quantification and regular PCR amplification using F3/B3 part of the RT-LAMP primer (Fig. S6A). The quantification of RT-LAMP is jointly determined by primer sets, and the influence of inner primers should also be considered. Our data suggest that the stronger the PCR using the F3/B3, the better the quantification of RT-LAMP (CDX2-p2, Ct = 25, Correlation coefficient = 0.99) in a serial dilution experiment. CDX2-p2 performed exceptionally well in all experiments compared to other primers. Amplification of DNA is visualized by GEL electrophoresis for both RT-PCR (Fig. S6B) and RT-LAMP (Fig. S6C) in control, colon cDNA sample and serial dilution experiment. Detailed analysis of a self-amplifying mouse Actb DNA-only LAMP primer set revealed that both FIP and BIP were the culprit (Fig. S6D). BIP self-amplified slowly, FIP self-amplified sharply after 25 min, whereas F3 + B3 did not self-amplify. A software algorithm to filter the FIP and BIP sequences can be carried out by searching for hairpin structures that enables self-amplification. RT-LAMP primers can be screened alongside the standard primer design process for stronger F3/B3 PCR and weak/no self-amplification to identify the best reagents for mRNA quantification. This strategy will make it convenient to identify a highly quantitative primer set. In the future, this process can be automated by developing appropriate machine learning models.
Conclusion
In summary, this work leveraged the advantages of the Loop Mediated Isothermal Amplification (LAMP) to develop a molecular diagnostic assay based on CDX2 expression. Cotton swab-based sample collection and direct sample testing procedures were developed, which were demonstrated to be superior to whole tissue chunks. The time from sample collection to CDX2 expression data was about 25 min. Optimization on the primer set could potentially reduce this time further. A 10–25-min diagnostic assay at the point of care setting will improve health care broadly across many diseases. This will also enable the development of consumer devices for at-home molecular testing. We hope this result will galvanize the community to develop innovative molecular testing both at home and at point of care settings.
Data availability
All data and materials are available in the main text or the supplementary materials. All codes used in the analysis are available in the github:sahoo00/LAMP.
References
Notomi, T. et al. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 28, E63 (2000).
Amaral, C. et al. A molecular test based on RT-LAMP for rapid, sensitive and inexpensive colorimetric detection of SARS-CoV-2 in clinical samples. Sci. Rep. 11, 16430 (2021).
Fellner, M. D. et al. Evaluation of RT-qPCR and loop-mediated isothermal amplification (LAMP) assays for the detection of SARS-CoV-2 in Argentina. Genes (Basel) 12, 659 (2021).
Lim, B. et al. Clinical validation of optimised RT-LAMP for the diagnosis of SARS-CoV-2 infection. Sci. Rep. 11, 16193 (2021).
Lu, S. et al. Multicenter international assessment of a SARS-CoV-2 RT-LAMP test for point of care clinical application. PLoS One 17, e0268340 (2022).
Marino, F. E., Proffitt, E., Joseph, E. & Manoharan, A. A rapid, specific, extraction-less, and cost-effective RT-LAMP test for the detection of SARS-CoV-2 in clinical specimens. PLoS One 17, e0266703 (2022).
Minami, K. et al. Evaluation of SARS-CoV-2 RNA quantification by RT-LAMP compared to RT-qPCR. J. Infect. Chemother. 27, 1068–1071 (2021).
Pu, R. et al. The screening value of RT-LAMP and RT-PCR in the diagnosis of COVID-19: Systematic review and meta-analysis. J. Virol. Methods 300, 114392 (2022).
A. Bektas et al., Accessible LAMP-Enabled Rapid Test (ALERT) for Detecting SARS-CoV-2. Viruses 13, (2021).
Ganguli, A. et al. Rapid isothermal amplification and portable detection system for SARS-CoV-2. Proc. Natl. Acad. Sci. U. S. A. 117, 22727–22735 (2020).
Inaba, M. et al. Diagnostic accuracy of LAMP versus PCR over the course of SARS-CoV-2 infection. Int. J. Infect. Dis. 107, 195–200 (2021).
K. R. Sreejith et al., A Portable Device for LAMP Based Detection of SARS-CoV-2. Micromachines (Basel) 12, (2021).
Li, Y. P. et al. Evaluation of RT-LAMP assay for rapid detection of SARS-CoV-2. Lab. Med. 54, 56–64 (2023).
Tapia-Sidas, D. A. et al. Starting from scratch: Step-by-step development of diagnostic tests for SARS-CoV-2 detection by RT-LAMP. PLoS One 18, e0279681 (2023).
Soroka, M., Wasowicz, B. & Rymaszewska, A. Loop-mediated isothermal amplification (LAMP): The better sibling of PCR?. Cells 10, 1931 (2021).
Damhorst, G. L. et al. Smartphone-imaged HIV-1 reverse-transcription loop-mediated isothermal amplification (RT-LAMP) on a chip from whole blood. Engineering (Beijing) 1, 324–335 (2015).
Ganguli, A. et al. Pixelated spatial gene expression analysis from tissue. Nat. Commun. 9, 202 (2018).
Ganguli, A. et al. Hands-free smartphone-based diagnostics for simultaneous detection of Zika, Chikungunya, and Dengue at point-of-care. Biomed. Microdevices 19, 73 (2017).
Wilisiani, F. et al. Development of a LAMP assay with a portable device for real-time detection of begomoviruses under field conditions. J. Virol. Methods 265, 71–76 (2019).
Enicks, D. A., Bomberger, R. A. & Amiri, A. Development of a portable LAMP assay for detection of neofabraea perennans in commercial apple fruit. Plant Dis. 104, 2346–2353 (2020).
Xun, G., Lane, S. T., Petrov, V. A., Pepa, B. E. & Zhao, H. A rapid, accurate, scalable, and portable testing system for COVID-19 diagnosis. Nat. Commun. 12, 2905 (2021).
Rafiq, A. et al. Development of a LAMP assay using a portable device for the real-time detection of cotton leaf curl disease in field conditions. Biol. Methods Protoc. 6, bpa010 (2021).
Parida, M., Sannarangaiah, S., Dash, P. K., Rao, P. V. & Morita, K. Loop mediated isothermal amplification (LAMP): A new generation of innovative gene amplification technique; perspectives in clinical diagnosis of infectious diseases. Rev. Med. Virol. 18, 407–421 (2008).
Marin, A. M. et al. Fluorescent and colorimetric RT-LAMP as a rapid and specific qualitative method for chronic myeloid leukemia diagnosis. Anal. Biochem. 641, 114541 (2022).
Deng, H. et al. An ultra-portable, self-contained point-of-care nucleic acid amplification test for diagnosis of active COVID-19 infection. Sci. Rep. 11, 15176 (2021).
Velders, A. H., Ossendrijver, M., Keijser, B. J. F. & Saggiomo, V. T-cup: A cheap, rapid, and simple home device for isothermal nucleic acid amplification. Glob. Chall. 6, 2100078 (2022).
Zou, M. et al. Rapid point-of-care testing for SARS-CoV-2 virus nucleic acid detection by an isothermal and nonenzymatic Signal amplification system coupled with a lateral flow immunoassay strip. Sens. Actuators B Chem. 342, 129899 (2021).
Choe, J. Y. et al. A comparative study of three different gene expression analysis methods. Technol. Health Care 25, 1073–1080 (2017).
Morishita, S. et al. Real-time reverse transcription loop-mediated isothermal amplification for rapid and simple quantification of WT1 mRNA. Clin. Biochem. 42, 515–520 (2009).
Nolan, T., Hands, R. E. & Bustin, S. A. Quantification of mRNA using real-time RT-PCR. Nat. Protoc. 1, 1559–1582 (2006).
Deepak, S. et al. Real-Time PCR: Revolutionizing detection and expression analysis of genes. Curr. Genom. 8, 234–251 (2007).
Bernardes, W. et al. Comparison of diagnostic performance of RT-qPCR, RT-LAMP and IgM/IgG rapid tests for detection of SARS-CoV-2 among healthcare workers in Brazil. J. Infect. Public Health 16, 1081–1088 (2023).
Beck, F. & Stringer, E. J. The role of Cdx genes in the gut and in axial development. Biochem. Soc. Trans. 38, 353–357 (2010).
Chawengsaksophak, K., James, R., Hammond, V. E., Kontgen, F. & Beck, F. Homeosis and intestinal tumours in Cdx2 mutant mice. Nature 386, 84–87 (1997).
Moskaluk, C. A. et al. Cdx2 protein expression in normal and malignant human tissues: An immunohistochemical survey using tissue microarrays. Mod. Pathol. 16, 913–919 (2003).
Drummond, F., Putt, W., Fox, M. & Edwards, Y. H. Cloning and chromosome assignment of the human CDX2 gene. Ann. Hum. Genet. 61, 393–400 (1997).
Mizoshita, T. et al. Expression of Cdx1 and Cdx2 mRNAs and relevance of this expression to differentiation in human gastrointestinal mucosa–with special emphasis on participation in intestinal metaplasia of the human stomach. Gastric Cancer 4, 185–191 (2001).
Bayrak, R., Haltas, H. & Yenidunya, S. The value of CDX2 and cytokeratins 7 and 20 expression in differentiating colorectal adenocarcinomas from extraintestinal gastrointestinal adenocarcinomas: Cytokeratin 7-/20+ phenotype is more specific than CDX2 antibody. Diagn. Pathol. 7, 9 (2012).
Huggett, J., Dheda, K., Bustin, S. & Zumla, A. Real-time RT-PCR normalisation; strategies and considerations. Genes Immun. 6, 279–284 (2005).
Cumberland, W. N. et al. Nonlinear calibration model choice between the four and five-parameter logistic models. J. Biopharm. Stat. 25, 972–983 (2015).
Spiess, A. N., Feig, C. & Ritz, C. Highly accurate sigmoidal fitting of real-time PCR data by introducing a parameter for asymmetry. BMC Bioinform. 9, 221 (2008).
Meagher, R. J., Priye, A., Light, Y. K., Huang, C. & Wang, E. Impact of primer dimers and self-amplifying hairpins on reverse transcription loop-mediated isothermal amplification detection of viral RNA. Analyst 143, 1924–1933 (2018).
Acknowledgements
SP had no funding support and worked as a volunteer. We thank Haddad lab members for helping during the experiment and mouse tissue collection. Images in the figures were created with BioRender.com.
Funding
National Institutes of Health grant R01-GM138385 (to DS).
Author information
Authors and Affiliations
Contributions
Conceptualization: D.S. Methodology: D.S., S.P., S.S.M., T.S. Investigation: D.S., S.P., S.S.M. Visualization: D.S., S.P. Funding acquisition: D.S. Project administration: D.S. Supervision: D.S. Writing – original draft: D.S. Writing – review & editing: D.S., S.P., T.S.
Corresponding author
Ethics declarations
Competing interests
Authors have disclosed the IP to UCSD. SP and DS are co-founders of the company Shanvi. SP is the President of Shanvi. All other authors have no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Pandey, S., McCoy, S.S., Stobdan, T. et al. Quantitative mRNA expression measurement at home. Sci Rep 14, 1013 (2024). https://doi.org/10.1038/s41598-023-49651-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-023-49651-8