Table 1 Precision assessment of hs-cTnI at three serum concentration levels.

From: Clinical performance validation and four diagnostic strategy assessments of high-sensitivity troponin I assays

Precision measurement parameter

Level 1

Level 2

Level 3

Hs-cTnI concentration mean(ng/L)

10.693

20.405

84.226

Intra-batch variance

0.546

1.402

4.091

Intra-batch standard deviation

0.739

1.184

2.03

Intra-batch coefficient of variation(CV)

6.91%

5.80%

2.40%

Inter-batch variance

0.222

1.130

6.433

Inter-batch standard deviation

0.471

1.063

2.536

Inter-batch coefficient of variation(CV)

7.16%

7.04%

3.59%